Unit is intended for adjustment and verification of:
- three-phase and single-phase meters of active, reactive or active and reactive energy;
- single-phase and three-phase wattmeters, varmeters and measuring transducers of active and reactive power;
- energy phase meters and frequency meters;
- voltmeters, ammeters and measuring converters of voltage and current in the industrial frequency range.
FEATURES:

Installation allows to carry out the following tests of meters:

- determination of relative error;

- determination of the standard deviation (S) when determining the error;

- checking the absence of self-propelled;

- starting current test;

- checking the meter constant;

- checking the counting mechanism;

- determination of errors when changing the phase sequence;

- determination of additional errors when changing the voltage and frequency of the network;

- determination of additional errors in the presence of harmonics in current and voltage circuits;

- determination of additional errors in case of load asymmetry;

- determination of additional errors in case of unbalance of phase voltages.


  • Availibility of a built-in stabilized voltage source with an output voltage regulation range from 1 to 300V

  • Availibility of a built-in stabilized current source with a range of regulation of the output current of each phase from 0.01 to 120A
  • Displaying the following results of measurements and calculations on indicator boards:

- RMS current values;

- root-mean-square voltage values;

- active, reactive and apparent powers in total and per phase in W, Var and VA, respectively;

-active and reactive energy in W*h, kW*h, var*h, kvar*h;

- errors of verified electric meters in%.

  • Software allows you to control the operation of the installation from a personal computer, which ensures the saving of verification results in databases with the possibility of generating verification results in the form of protocols. Software allows the user to independently create the type of verification protocol

No

Parameter

Unit

Parameter value

1

Current regulation range

A

3×(0,01ナ120)

2

voltage regulation range

(Unom =3�220 V)

V

3×(1ナ300)

3

Output power, per phase:

In current circuit

In current circuit

V��

100

50

4

Non-sinusoidality of current and voltage signals at

maximum load, no more than

%

0,5

5

Phase control range

Phase angle setting accuracy

degrees

0ナ360

ᄆ0,1

6

Frequency control range

Frequency setting accuracy

Hz

45-65

0,05

7

Defined harmonics

2ナ31

8

Installation error:

-current

-voltage

%

ᄆ 0,5

ᄆ 0,5

9

Instability of set values for 120 s of:

-current

-voltage

-ᅠpower

%

ᄆ 0,05

ᄆ 0,05

ᄆ 0,05

10

Limits of permissible relative error of the reference meter HY-5303C, in the ranges of

voltages 3×(40ナ480)V

currents 3×(0,01ナ120)A

when measuring active energy, no more than

when measuring reactive energy, no more than

%

Accuracy class

0,05

0,1

ᄆ 0,05

ᄆ 0,1

ᄆ 0,1

ᄆ 0,2

11

Number of simultaneously checked meters

pcs.

1ナ3

12

Operating temperature range

º�

23 ᄆ 3

13

Supply voltage

V

3×220/240 ᄆ 10%

14

Mains frequency

Hz

50 or 60

15

Power consumption, no more than

V∙�

500

16

Overall dimensions (length; width; height)

mm

800 � 600 � 1340

17

Weight

kg

110