Automatic installations for verification of electricity meters

NEVA-Test 9303 reference meter can be used independently, in combination with a PC that expands its functionality, as well as as part of specialized and universal verification installations.

NEVA-Test 9303 reference meter is recommended for use in metrological laboratories of large industrial enterprises, power systems and CSM.

The reference meter NEVA-Test 9303 is a wide-range multifunctional electricity meter of accuracy class 0.02 and is intended for calibration and verification of the following reference and working instruments for measuring electric power quantities:

- single-phase and three-phase meters of active and reactive electrical energy;

- single-phase and three-phase wattmeters, varmeters and measuring transducers of active and reactive power;

- phase meters and frequency meters;

- electrical measuring instruments (voltmeters, ammeters) and measuring voltage and current converters in the industrial frequency range;

- means for measuring and recording the parameters of electrical energy in single-phase and three-phase electrical networks.

FEATURES:

- pulse inputs and outputs have a convenient location for use on the front and rear panels;

- error calculation can be carried out simultaneously on 4 pulse inputs;

- the value of the pulse output constant can be set by the user;

- touch screen for control, convenient and intuitive interface;

- external mouse and keyboard can be used for control;

- Availability of communication ports RS-232, USB and Ethernet.

Metrological characteristics

Parameter

Range

Error

Phase voltage

10 ~ 600V

ᄆ 0,02 %

three ranges
40V, 400V, 600V

Current

1mA ~ 120A

ᄆ 0,02 %

five ranges
10m�, 100m�, 1�, 10�, 120�

Current and voltage harmonics

2 to 59

Frequency

45 ~ 65 Hz

ᄆ 0,001 Hz

Active, reactive and apparent power and energy

ᄆ 0,02 %

Phase angle

from 0 up to360

� 0,005

Specifications

Parameter

Value

Input frequency of pulse inputs

up to 1 MHz

Output frequency of pulse outputs

up to 1 MHz (pulse duration> 0.5 ᄉs)

Output constant:

frequency of the output pulses in automatic mode is 60 kHz at nominal values

Power consumption

<50V￀

Dimensions

482 × 139 × 496 mm

weight

11.5 kg

Environmental conditions:

Temperature +15 �� - +40 ��

Relative humidity 40% - 75%

Quality.

        Reliability.